ASTM F – 11a Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants During Magnetic Resonance. ASTM Fa. Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants During Magnetic. ◇ASTM Fa for Measurement of. Measurement of Radio Frequency Induced. Heating Near Passive Implants During MRI. ◇ASTM.

Author: Dujas Duzahn
Country: Eritrea
Language: English (Spanish)
Genre: Technology
Published (Last): 23 December 2006
Pages: 181
PDF File Size: 9.44 Mb
ePub File Size: 6.6 Mb
ISBN: 556-6-15183-506-6
Downloads: 64196
Price: Free* [*Free Regsitration Required]
Uploader: Gole

Referenced Documents purchase separately The documents listed below are referenced within the subject standard but are not provided as part of the standard. The RF coil is assumed to have quadrature excitation.

ASTM International – Standard References for ASTM F – 11a

No other units of measurement are included in this standard. For other implantation conditions for example, external fixation devices, percutaneous needles, catheters or tethered devices such as ablation probesmodifications of this test method are necessary.

Historical Version s – view previous versions of standard. The heating measurements are made twice, once with the implant and then repeated at the same location without the implant.

This test method describes a test procedure for evaluating the RF-induced temperature rise associated with an MR procedure involving a specific frequency of RF irradiation of an implant.

  3M 468MP PDF

Link to Active This link will always route to the current Active version of the standard. These two measurements estimate the local SAR and the local additional temperature rise with xstm implant.

Active view current version astk standard. Work Item s – proposed revisions of this standard. No other units of measurement are included in this standard. While the focus in this test method is on 1.

Other safety issues asttm should be addressed include magnetically induced displacement force and torque, as well as proper device function while in various configurations in the MR environment.

For other implantation conditions for example, external fixation devices, percutaneous needles, catheters or tethered devices such as ablation satmmodifications aztm this test method are necessary.

The RF coil is assumed to have quadrature excitation. The combination of the test results and the computational model results may then be used to help assess the safety of a patient with the implant during an MR scan. Work Item s – proposed revisions of this standard. Other safety issues that should be addressed include magnetically induced displacement force and torque.

It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. While the focus in this test method is on 1.

  ELECTROLUX ENN 3153 AOW PDF

ASTM F2182 – 11a

Link to Active This link will always route to the current Active version of the standard. Referenced Ashm purchase separately The documents listed below are referenced within the subject standard but are not provided as part of the standard. The results may be used as an input to a computational model for estimating temperature rise due to the presence of asym implant in a patient. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Because of possible additional heating, particularly when implant dimensions approaches or exceeds one quarter of the wavelength of the RF field inside the phantom, conclusions from measurements made at one static magnetic field strength do not apply to other field strengths asgm frequencies.